Control button
jams, moves
erratically.
Piston contaminated.
Seals contaminated.
Penetration of solvent
vapors.
Clean piston and lubricate
lightly.
Disassemble pipette. Clean
all seals and exchange
where necessary
(see Part B, Maintenance).
Unscrew pipette at the
central junction and
ventilate.
Clean piston and lubricate
lightly.
Pipette
blocked,
too little liquid is
aspirated.
Liquid has penetrated the tip
cone and dried.
For 25 to 500 μL pipettes:
The filling tube in the nose
cone is blocked.
Unscrew pipette at the
central junction, rinse lower
part first with warm water,
then with distilled water and
allow to dry.
Or:
Disassemble pipette.
Replace ejector seal in the
tip cone (see Part B, Mainte-
nance).
For 25 to 50 μL:
Replace tip cone.
For 100 to 500 μL:
Replace the filling tube in
the tip cone (see Part B,
Maintenance).
Error Cause Solution
Eppendorf Reference® – Part A – Chapter 10